Model 10A16-4C
Quad Platinum RTD Conditioner Card
|
Input Type
|
Description
|
SPS6000
|
SPS8000
|
System 10
|
|
Temperature
|
Quad Platinum RTD
|
X-X
|
X X
|
X X
|
This conditioner is for high-accuracy temperature measurement using platinum
Resistance Temperature Detectors (RTD's), either DIN (European) or American
standard. Precision constant-current excitation is provided for four independent sensor
channels, which may be intermixed as required. RTD inputs are normally set for four-wire
cabling, with maximum excitation of one milliampere and input impedance of 10
megohms, to eliminate common self-heating and cable-loading errors. However, for an
RTD channel with shared current return and sense signals, three-wire mode is available,
if desired, via an internal jumper setting.
Calibration of the 10A16-4C may be either
- "absolute" for RTD's of DIN standard (only), or
- "two-point" (i.e., "zero and span"), when independently and accurately known temperature references are available- preferably the high and low extremes to which the sensor will be subjected.
Specifications:
- RTD Types: Platinum; DIN (European) standard with "ice-point" resistance of 50, 100, or 200 ohm, or American standard with "ice-point" resistance of 100 ohm
- Range and Resolution: See Table below; automatically selected- on an individual channel basis- when the channel is configured
- Linearization: Internal digital linearization; maximum error: ±0.1° C (±0.2° F), System 10 and SPS8000
- Excitation (per channel): Nominal 5 V-DC; 1 mA, maximum
- Amplifier (per channel):
Input Impedance (Differential): 10 M-ohm
Offset: Initial: ±0.1 mV; vs. Temperature: ±0.5 µV/°C; vs. Time: ±5 µV/month
Gain Accuracy: ±0.02% of full scale
Gain Stability: vs. Temperature: ±25 ppm/°C; vs. Time: ±20 ppm/month
- Filter (per channel): 3-pole modified Butterworth; 3 dB down at 2 Hz; 60 dB down at 25 Hz
Step-Response Settling Time (Full-Scale Output):
To 1% of final value: 0.6 sec
To 0.1% of final value: 0.8 sec
To 0.02% of final value: 1.5 sec
- Total System Accuracy (typical, including Model 10A16-4C and system data collection and processing): see Table below
- Auxiliary Outputs: Filtered outputs available on wire-wrap pins
RTD Ranges for the Models 10A16-4C
|
Platinum RTD Standard
|
Resistance
|
Range
|
Display Resolution3
|
Typical System Accuracy4
|
Max. Expected Error4
|
DIN1
DIN1
DIN1
DIN1
|
50 ohm
50 ohm
50 ohm
50 ohm
|
-150°C to +600°C
-150°C to +600°C
-238°F to +1112°F
-238°F to +1112°F
|
1.0°C
0.1°C
1.0°F
0.2°F
|
±1°C
±0.2°C
±1°F
±0.3°F
|
±1.3°C
±0.4°C
±1.5°F
±0.6°F
|
DIN1
DIN1
DIN1
DIN1
|
100 ohm
100 ohm
100 ohm
100 ohm
|
-150°C to +600°C
-150°C to +600°C
-238°F to +1112°F
-238°F to +1112°F
|
1.0°C
0.1°C
1.0°F
0.2°F
|
±1°C
±0.2°C
±1°F
±0.3°F
|
±1.3°C
±0.4°C
±1.5°F
±0.6°F
|
DIN1
DIN1
DIN1
DIN1
|
200 ohm
200 ohm
200 ohm
200 ohm
|
-150°C to +600°C
-150°C to +600°C
-238°F to +1112°F
-238°F to +1112°F
|
1.0°C
0.1°C
1.0°F
0.2°F
|
±1°C
±0.2°C
±1°F
±0.3°F
|
±1.3°C
±0.4°C
±1.5°F
±0.6°F
|
American2
American2
|
100 ohm
100 ohm
|
-150°C to +600°C
-238°F to +1112°F
|
0.1°C
0.2°F
|
±0.2°C
±0.3°F
|
±0.4°C
±0.6°F
|
NOTE: Ranges printed in BOLDFACE apply to System 10 only.
1 alpha = 0.00385 (Back to Table)
2 alpha = 0.00392 (Back to Table)
3 For the Model 10A16-4C, applies to System 10 only. Note that a specially modified version of the 10A16-4C yields resolution of 0.01° C (0.02° F) when used with System 10 "custom linearization." (Back to Table)
4 Including ± one count of least significant digit displayed (i.e., ± 1, 0.2, or 0.1) (Back to Table)
|
|