The
single-channel Models 5D78 and 5D78V are similar to the DC-excited Models
5D70 and 5D70V, respectively, but feature a highly effective phase-sensitive
carrier-amplifier design. Three selectable operating frequencies are
provided to best match the user's specific transducer.
The
5D78 delivers filtered analog output of ±5 VDC, while the 5D78V
produces ±10 VDC.
The
5D78(V)'s advanced analog design directly addresses the problem of
measurement inaccuracy in industrial environments of high
electromechanical noise. Exceptional signal stability and
accuracy over the entire input range are achieved through
 |
regulated,
remotely sensed AC excitation |
 |
precise
LINEARITY / SYMMETRY correction and SIGNAL PHASE adjustment via
software commands |
 |
high-stability
amplification |
 |
configurable
low-pass active filtering |
 |
"absolute"
and "two-point" software-based calibration |
 |
effective
signal isolation and ESD protection |
While
intended primarily for applications involving transformer-coupling to the
transducer bridge (as with conventional rotary-transformer torque
sensors), the Model 5D78(V) can also be used when high sensitivity is
required or where the electrical environment is especially noisy.
Responding only to the modulated carrier frequency, the module rejects
extraneous voltages that can cause errors in DC systems, particularly when
there is a need to "blow up" a portion of the transducer range.
For both the Model 5D78V and the Model 5D78S (the "S"
Option version of the Model 5D78), a 10-pin header for a Phoenix
Style screw-terminal connector
replaces the Model 5D78's standard DB25 connector for the module's power,
configuration, logic, and analog output wiring. Ensuring secure cable connections to individual 5D modules,
these versions are intended primarily for single-point or
single-mount module applications. Since the logic "enable"
line is not provided in this case, both analog outputs are continuously
available for both the 5D78V and the 5D78S.
The
5D78/5D78V Overcomes Errors That Traditionally Plague the Strain-Gage
Measurement Process
For
steady indication and smooth, dependable control action, the 5D78(V) can
provide a true average value of the measured variable, even in the face of
substantial dynamic content. As with most 5D Series modules, two
high-level, noise-free analog output signals are provided. For the
5D78 (only), one output may be switched on and off by setting the logic
state of an "enable" line. For the 5D78V, both outputs are
continuously available.
Both
models feature
 |
Powerful
low-pass active filtering, independently selectable for each
output, for removal of unwanted high-frequency measurement-signal
components and the elimination of aliasing errors, if the module's
output is subsequently sampled |
 |
High
noise rejection, eliminating errors from common-mode pickup and
ground-loop coupling, with 1500 VAC isolation between input and
output terminals and between I/O and power supply / communications
terminals |
 |
High
ESD immunity and extensive EMI protection further assure data
integrity in harsh industrial environments |
Internal
"Absolute" Calibration Coupled with High-Accuracy
"Two-Point (Deadweight)" Calibration
Because
of normal cable loading effects, it is a practical necessity to calibrate
any AC TORQUE SENSOR/CABLE/INSTRUMENT system after installation, using a known
input standard. The "absolute" calibration provided by
the 5D78(V) normally yields a very good first approximation.
Conventional "zero and span" calibration may then be easily
performed—with the module fully "in place"—using the
Configurator software supplied with the module. This procedure
includes any necessary phase-shift compensation, symmetry trimming, and
midscale linearity adjustments.
When
calibrated "in place," the 5D78(V) offers excellent stability and
interchangeability of units. As long as an initially calibrated
5D78(V)'s setup configuration is exactly transferred to the 5D78(V) that
replaces it, no recalibration is usually required to maintain the
stated accuracy (see Specifications).
The
module also allows fast and easy software-controlled calibration by means
of a user-installed shunt resistor.
Model
5D78(V) Specifications [Top
]
(Click
here for 5D Series General Specifications)
|
|
Transducer
Types |
Conventional 4-arm
strain gage bridges, typically transformer-coupled, 120 Ω to 10 kΩ; zero range is 20% of the
stated full scale. A screw terminal is provided for user-supplied
shunt calibration resistor. The diagrams below show standard 4-wire
cabling (for under 20 ft. in length) and 8-wire cabling (for 20 ft. or
longer).
|
|
|
|
Input
Ranges (Nominal, Full-Scale) and Accuracy per Excitation Setting
|
See
table, below; selectable when the 5D78(V) is configured. Accuracy
is given as % of full-scale overall expected maximum error, following
calibration. Note that the highest range selection accommodates
actual inputs as high as 4.8 mV/V.
|
|
Excitation |
| Range
(mV/V) |
3.27
or 5.00 kHz |
10.00
kHz |
| 0.5 |
0.04 |
0.06 |
| 0.75 |
0.03 |
0.04 |
| 1 |
0.02 |
0.03 |
| 1.5 |
0.02 |
0.02 |
| 2 |
0.02 |
0.02 |
| 3 |
0.02 |
0.02 |
|
|
Excitation |
|
Frequency |
3.27,
5.00, or 10.00 kHz; selectable when the 5D78(V) is
configured |
|
Voltage |
Nominal
3 VAC rms |
|
Accuracy |
Dependent on range
and excitation; see table, above
NOTE: The full-scale accuracies given in the table refer to the
5D78(V)'s response to substantially undistorted waveform inputs.
Any phase shift must be compensated for during calibration. Some
gain deviation from transducer data-sheet specifications is normally to be
expected as a result of loading the sensor's finite output impedance
with intrinsic cable capacitance. The degree of this error will vary
with cable type and length. |
|
Amplifier |
|
Normal-Mode
Range |
15
V rms operating; 240 V rms without damage |
|
Input
Impedance (Differential) |
Greater than
350 kΩ |
|
Offset |
Initial:
±0.05% of full scale
Vs. temperature: ±25 ppm/°C
Vs. time: ±10 ppm/month |
|
Gain
Accuracy |
±0.02% of full
scale typical, following calibration; see table, above |
|
Gain
Stability |
Vs. temperature:
±25 ppm/°C;
Vs. time: ±10 ppm/month |
|
Analog
Filters |
0.2,
2, 20, 200, or 2000 Hz, independently selectable for each output
NOTE:
For the Model 5D78(V), the value of the highest allowed corner frequency is dependent
on the selected excitation frequency: for an excitation of 10.00 kHz, it
is 2000 Hz; for 5.00 kHz, it is 1400 Hz; and for 3.27 kHz, it is 1100 Hz. |
|
Analog
Outputs |
Filtered
± 0 to 5 VDC (for the Models 5D78 and 5D78S) or ± 0 to 10 VDC
(for the Model 5D78V), with linearity maintained for 20% overrange.
For the Model 5D78 (only), Output B is switchable via logic
"Enable" line. |
|
Logic-Input
Functions
(see
also General 5D Specs) |
Enable, Not
Calibrate A, Not Calibrate B
The following
diagrams show 5D78 logic input connections for switch closure (no
external supply) and active TTL logic, respectively, using the standard
module DB25 connector. For the
"Enable" function, see 5D Analog
Outputs and the description of the MODULE
IDENTIFICATION (MID) command. NOTE:
For the Models 5D78V and 5D78S, the Enable
function is not provided, and both analog outputs are continuously
available. |
|
|
|