Analog Input Instruments

Models 3163, 3263, 3363

These instruments accept inputs from various analog voltage sources, including DC-to-DC LVDT's, potentiometer-type sensors, Hall-Effect devices, photocells, current shunts, and the outputs of other instrument systems with various grounding configurations and voltage/impedance levels.  

The Model 3163 Analog Input Conditioner is the basic "Form 1" instrument.

The Model 3263 Analog Input Conditioner / Indicator is the "Form 2" instrument, providing vivid front-panel digital indication of measured values, scalable in desired engineering units.

The Model 3363 Analog Input Conditioner / Indicator / Controller is the "Form 3" instrument, and includes HI/LO limit detection with control output.

The signal source can be grounded or floating, and may use 2-, 3-, or 4-wire cabling, as shown below.  A regulated 10-V supply is provided for excitation of potentiometers, DC-to-DC sensors, and similar devices.  Full-scale sensitivity is continuously adjustable in four jumper-selectable ranges.  Three standard 5-V outputs are provided (see Specifications).

Signal Source Configurations for the Models 3163, 3263, 3363

With a sensitive, high-impedance floating differential input, wide common-mode range, and exceptional common-mode rejection, these models can be used to obtain difficult low-level signals from highly off-ground sources, without common-mode AC or DC offset problems.

Selectable low-pass active filtering allows the smoothing of unwanted normal-mode dynamic components that might prevent stable digital conversion or control action.

For a typical 3263 operation, see the following application:

3-Phase AC Power Measurement

3000 Series options applying to these instruments include

Analog Peak Capture (Models 3263 and 3363)

4-20 mA Current Output (all models)

0-10 VDC Dual Galvanic Isolated Output (Models 3163 and 3263)

Internal Relays, Electromechanical and Solid-State (Model 3363)

12 VDC Battery-Powered Operation (all models)

Nominal 230 VAC Operation (all models)


Analog Input Instrument Specifications  [Top ]

(Click here for 3000 Series General Specifications)

Input Type

2-, 3-, or 4-wire DC voltage source, floating or grounded (see diagram, above) 

Input Ranges (full-scale)

Full-scale sensitivity continuously adjustable in jumper-selectable ranges of 50 to 500 mV, 500 mV to 5 V, 5 to 50 V, and 50 to 250 V

Excitation Supplied

Regulated 10 VDC ±0.02%, 20 mA maximum, for excitation of potentiometers, DC-to-DC sensors, and similar devices

Analog Outputs

Three outputs, each ±5 VDC full-scale with 50% overrange, 5 mA maximum; low-pass corner frequencies of 2 Hz, 200 Hz, and 2 kHz, respectively

Common-Mode Range

±100 VDC

Common-Mode Rejection Ratio

-70 dB, DC to 60 Hz

Input Impedance

1 MΩ (all ranges)

Analog Filtering

Active low-pass filters provide -60 dB per decade above cutoff frequency ("f"); full-scale slew time is 1.4/f sec

Output Ripple and Noise

0.15% of full scale (rms) maximum for 200-Hz and 2-kHz outputs; 0.02% of full scale (rms) maximum for 2-Hz output

Accuracy (typical, following calibration)

0.1% of full scale for 60 days ("hands off") following initial calibration

Display Resolution (Models 3263 and 3363)

0.02% of full scale