With
the installation of an optional circuit board, real-time analog peak
capture is possible for a Form 2
or Form 3 instrument, in addition to
its normal "tracking" function.
Controlled
either by front-panel push buttons or by a remote user-installed switch, a
"P" unit can "freeze" and display the last positive
input-signal "peak" value to have been perceived. The
analog output is held at this value until a RESET command is
applied. A subsequent higher signal excursion disables the
"hold," permitting the capture of subsequent higher peaks.
To
prevent low-signal noise from triggering a "peak hold," the peak
capture function is automatically disabled when the input is less than 8%
of full scale. If you require peak capture within the 0-8% range,
contact the factory.
Because
it is digitally held, a displayed peak value will not decay. The
maximum decay rate for a held analog output is only 0.1% of full scale per
minute.
With
all Form 3 instruments, limits are
automatically latched upon peak capture, and are automatically reset on
return to normal "track" mode.
For
a special "P" version dedicated to the capture and hold of peak
torque values, see the Model TM3 Torque
Monitor.
NOTE:
The "P" Option may NOT be used in combination with the
"C," "G," "R," or "S" Option.
|

|
|