With
this option, a Form 1 or Form
2 instrument can furnish two independent galvanic outputs, each 0-10
VDC (2 mA maximum), full scale. These outputs are completely
isolated not only from each other but also from the 3000 instrument's
"common."
The
use of galvanically isolated outputs prevents ground-loop effects in
interconnections with remote data-acquisition systems or
controllers. The presence of two independent outputs lets you send
collected data to two different systems or devices, each with its own
ground.
Load
limit for each output exceeds 10 kΩ. Output bandwidth is
normally 40 Hz; the "G" option can be easily modified, however,
for other bandwidths up to 500 Hz.
NOTE:
The "G" Option may NOT be used in combination with the
"C," "P," "R," or "S" Option.
 |
|
Isolated
Voltage Output to Strip Chart Recorder
|
"G"
Option Specifications [Top
] |
|
Output
Range
|
±10
V-DC full scale (2 mA maximum), normal; internal controls give
approximately ±5% of adjustment authority on both SPAN and ZERO
|
|
Common-Mode
Range |
±500
V, maximum |
|
Common-Mode
Rejection Ratio |
-120
dB at DC; -60 dB at 60 Hz |
|
Gain |
Factory
set to 2.00; user-adjustable within ±5% |
|
Linearity
|
±0.1%
of full scale
|
|
Maximum
Zero or Span Drift (after warmup to 1/2 hour)
|
±0.2%
of full scale*
|
|
*
Applies to the "G" Option only, and does not include
possible drift contributed by the signal conditioner board of the
base 3000 instrument. |
|