Used
as a DC ammeter, this Digital Panel
Meter (DPM) model provides a full-scale readout of ±20,000 counts
(selectable fixed zero or active least significant digit) and four
full-scale direct-reading current ranges/resolutions (see Specifications).
The 5.000-A range measures the
IR drop across a built-in 10-mΩ
current shunt. Accuracy is 99.99% of full scale ±
1 count.
The
model-numbering system is given in the Specifications
table, below.
All
current ranges are calibrated at the factory, with calibration factors for
each range stored in internal EEPROM.
This allows ranges to be changed in the field without having to
recalibrate the meter.
As
with all 2000 Series DPM’s, the peak value of the input signal is
automatically captured and may be displayed
via front-panel push button or a control signal received at the rear
connector (since the last meter reset). Data readings and outputs can be separately selected to be
either filtered or unfiltered.
See Digital Panel Meters (DPM’s)
for a full description of these and other standard DPM features.
2000
Series options applying to the DC current meter include
DC
Current Meter Specifications [Top
]
(Click
here for 2000 Series General Specifications)
|
|
Model
Number
|
DC
Current
Range
|
Resolution
|
Input
Ohms
|
Error
± 1 Count
|
|
2161-2
|
2.0000 mA
|
0.1 μA
|
100 Ω
|
0.01 % full scale
|
|
2161-20
|
20.000 mA
|
1 μA
|
10 Ω
|
0.01 % full scale
|
|
2161-200
|
200.00 mA
|
10 μA
|
1 Ω
|
0.01 % full scale
|
|
2161-5
|
5.000 A
|
1 mA
|
0.01 Ω
|
0.04 % full scale
|
|
All DC
Current Models
|
|
Span
Temperature Coefficient |
±0.003% of
reading/°C |
|
Zero
Temperature Coefficient |
0.1 count/°C |
|
Normal-Mode
Rejection at 50/60 Hz |
90 dB with minimum
digital filtering |
|
Common-Mode
Rejection from DC to 60 Hz |
130 dB |
|
Overcurrent
Protection |
25x for 2 mA, 8x
for 20 mA, 2.5x for 200 mA, 1x for 5 A |
|