These
Digital Panel Meter (DPM) models serve as high-sensitivity monitors and
controllers for use with load cells, strain gages, and other
microvolt
input signals where high accuracy and stability are required.
See, for example, the following
applications:
The
Model 2170 uses standard six-wire connection to load cells, where the
sense lines are separate from the excitation lines. (Standard four-wire
cabling, where the excitation and sense lines are tied together, can also
be accommodated.) By eliminating effects due to variations in lead resistance, this
configuration allows long cable runs in outdoor environments with
temperature extremes.
The
"X" version (Model 2170X) allows
custom scaling, as specified by
the user at the time of order (specify minimum input value, minimum
displayed reading, maximum input value, and maximum displayed reading).
Contact the
factory for more details.
When
used directly as a DC microvoltmeter, either of these instruments can
provide sensitivity down to 20 mV full-scale with 1 µV resolution.
With a digital multiplier of five, 99,999 counts (selectable fixed
zero or active least significant digit) can be displayed
with a sensitivity of 0.2 µV per count. (In this case, a moving
average filter with a time constant of 1.2 sec or greater should be
selected.)
When
used as a load cell meter, the 2170 or 2170X allows scaling for direct
readout in pounds, kilograms, PSI, or other desired engineering units. Scaling can be via front-panel push buttons or computer
commands. Zero may be set
from -99,999 to +99,999, and range (i.e., span) may be scaled from 0 to ±99,999.
Digital scaling and calibration eliminate zero and span drift
associated with potentiometers in non-microcomputer based meters.
An isolated 10
VDC
excitation supply can provide up to 120 mA of current to power up to four
350-Ω load cells in parallel (the excitation and sense points of the
four bridges are connected in parallel through a summing box).
The meter can be set to operate in a ratiometric mode to eliminate
errors due to supply variations. With
standard six-wire connections, the 2170(X) also compensates for variation
in resistance of the transducer leads, thereby allowing long cable runs.
The
strain gage meters allow easy setting of scale and offset through either
the coordinate reading method or the manual coordinate method (explained
under Digital Panel Meters (DPM’s)).
The
auto-tare function provided by all 2000 Series DPM's is very useful when
the strain gage meters are used in weighing applications, as in the first
two applications listed above. Here,
the displayed value is set to zero to subtract out the weight of an empty
container. This function may
be controlled by an external push-button contact closure or by a logic
signal to the rear of the meter
As
with all 2000 Series DPM’s, the peak value of the input signal is
automatically captured and may be displayed
via front-panel push button or a control signal received at the rear
connector (since the last meter reset). For a typical peak capture application, see Using
Peak Capture to Test Wire Tensile Strength. Data readings and outputs can be separately selected to be
either filtered or unfiltered.
See Digital Panel Meters (DPM’s)
for a full description of these and other standard DPM features.
2000
Series options applying to the RMS voltage meter include
Strain
Gage Meter Specifications [Top
]
(Click
here for 2000 Series General Specifications)
|
|
DC
MICROVOLTMETER Inputs
|
|
Input
Range (mV)
|
Resolution
|
Error (at 25°C,
for all ranges)
|
|
±20.000
|
1.0 μV
|
0.1% of full scale ± 1
count
|
|
±50.000
|
2.5 μV
|
|
±100.00
|
5.0 μV
|
|
±250.00
|
12.5 μV
|
|
±500.00
|
25 μV
|
|
LOAD
CELL Inputs
|
|
Full-Scale
Input (mV)
|
Zero
Adjust
(all ranges)
|
Span
Adjust
(all ranges)
|
Error (at 25°C,
for all ranges)
|
|
±20.000
|
-99.999 to +99,999
|
0 to ±99,999
|
0.1% of full scale ± 1
count
|
|
±50.000
|
|
±100.00
|
|
±250.00
|
|
±500.00
|
|
Span
Temperature Coefficient |
±0.0015% of
reading/°C |
|
Zero
Temperature Coefficient |
±0.1 μV/°C |
|
Normal-Mode
Rejection at 50/60 Hz |
90 dB with minimum
digital filtering |
|
Common-Mode
Rejection from DC to 60 Hz |
130 dB |
|