ILA Series General Specifications

Housing

In-Line mount housing; non-removable screw terminals. Plastic ABS

Dimensions

Connections

The ILA I/O CONNECTIONS are internal - via non-removable screw terminals which will accept wire sizes from AWG 16 to 30. NOTE: The recommended transducer cabling using sense wires would be eight wire, individually shielded, twisted pair - wired as indicated per modules data sheet.

Power

11- 28 Vdc ; 2 watts Max, typical

Input Overvoltage Protection

Up to 240 VAC rms on all Signal and Excitation lines

Conformity   

Directives: 2014/30/EU Electromagnetic Compatibility

                    2014/35/EU Low Voltage Supply

Analog Filtering

Low pass modified 3-pole Butterworth - selectable low, medium & high settings

Operating Temperature Range

-10° C to 70° C (14° F to 158° F)

Operating Relative Humidity

5% to 95%, noncondensing

Analog Outputs

Filtered ± 0 to 5 Vdc or ±0 to 10 Vdc, 4-12-20 or 4-20 mA (sourcing). Mode is switch selectable with linearity maintained for 20% over-range (in voltage mode only)

 Accuracy

Offset and Gain Stability ±30 ppm/Deg C; vs. time ±10ppm/month: Gain Accuracy is limited to calibration source

Logic Inputs

Activated with power common to initiate functions such as NULL, SHUNT. Specific to Model. Internal utility switch provided for local activation

Status Indicator Light

GREEN

Power supply indicator

YELLOW

Input signal over 20% out of range

 

 
 
 