MPUT

MPUT

 

MODEL MPUT UNIVERSAL THERMOCOUPLE POD

 

 

MPSTR

MPSTR

 

MODEL MPSTR DC Strain Gage POD

 

 

MPUC

MPUC

 

MODEL MPUC UNIVERSAL CURRENT CLAMP POD

 

 

(for use with the “MP-TR” Current Clamp Probes listed on the next page)
Operating Range (Differential): 1.5 VAC RMS = Rated full scale of probe;
Crest Factor: 2
Resolution: Dependent on clamp probe
Accuracy: Dependent on clamp probe (typically 1 to 3 % for inputs greater
than 10% of full scale)
Connection: Hypertronics connectors for “MP-TR” Probes

MP600V

MP600V

 

MODEL MP600V VOLTAGE POD

 

 

Operating Range (Differential): ±0 to 600 VDC or 0 to 600 VAC RMS;
will sustain ±1000 VDC for one minute without instrument damage
Common-Mode Range: ±600 VDC operating and without instrument
damage
Resolution: Tenths of a volt
Input Impedance: 16 MΩ (differential); 8 MΩ (common-mode)
Accuracy: ±0.1% of full scale up to 400 Hz; ±0.5% of full scale up to 3 kHz;
±1% of full scale up to 6 kHz
Connection: Safety banana jacks

MP30V

MP30V

 

MODEL MP30V VOLTAGE POD

 

 

Operating Range (Differential): ±0 to 30 VDC or 0 to 24 VAC RMS;
will sustain ±400 VDC for one minute without instrument damage
Common-Mode Range: ±30 VDC operating; ±200 VDC without
instrument damage
Resolution: Millivolts
Input Impedance: 4 MΩ (differential); 1 MΩ (common-mode)
Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full scale up to 5 kHz
Connection: Five-way binding posts

MP300V

MP300V

 

MODEL MP300V VOLTAGE POD

 

 

Operating Range (Differential): ±0 to 300 VDC or 0 to 240 VAC RMS;
will sustain ±600 VDC for one minute without instrument damage
Common-Mode Range: ±300 VDC operating and without instrument
damage
Resolution: Hundredths of a volt
Input Impedance: 40 MΩ (differential); 10 MΩ (common-mode)
Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full scale up to 5 kHz
Connection: Safety banana jacks

MP20MA

MP20MA

 

MODEL MP20MA CURRENT POD

 

 

Operating Range (Differential): ±0 to 20 mADC or 0 to 14 mAAC RMS;
will sustain ±50 mADC for one minute without instrument damage
Resolution: Microamps
Input Impedance: 250 Ω (differential); 500 kΩ (common-mode)
Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full scale up to 5 kHz
Connection: Five-way binding posts